How to prevent the ingress of moisture through the cable of submersible pressure transmitters and level probes?

If you wish to measure the level of a liquid easily and reliably, a lot of people will do this using hydrostatic pressure measurement, e.g. with a submersible pressure transmitter or perhaps a so called level probe. The characteristic submersed application implicates a maximum contact with the surrounding, mainly water-based medium, respectively to ?moisture?.
Exposure isn’t only limited by the wetted elements of the pressure sensor housing, but also to the entire immersed length of the cable. In addition, beyond your directly immersed level probe parts, the cable, and specifically the cable end, tend to be exposed to moisture due to splash water, rain and condensation. This is true not merely during operation, but a lot more during installation and commissioning, or when maintenance or retrofitting is required. Irrespective of the mark application, whether in water and wastewater treatment or in tank monitoring, moisture ingress in to the cable ends of the submersible pressure transmitter may appear early and irreversibly with insufficient protection measures, and, in virtually all cases, lead to premature failure of the instrument.
The ingress of moisture into the cable outlet and from there on downwards in to the electronics of the level probe must be actively eliminated by preventive actions by an individual. To gauge the level with highest accuracy, the varying ambient pressure above the liquid media, which is also ?resting? on the liquid, should be compensated against the hydrostatic pressure functioning on the pressure sensor (see article: hydrostatic level measurement).
Ventilation tube
Thus, it really is logical that there surely is a constant risk of a moisture-related failure because of moisture ingress (both via the ventilation tube and through the actual cable itself) if you can find no adequate precautionary measures. To pay Festive ?resting? on the media, a ventilation tube runs from the sensor element within the particular level probe, through the cable and from the level probe by the end of the cable. Due to capillary action within the ventialation tube used for ambient pressure compensation, moisture can be transported from the surrounding ambience right down to the sensor.
Thus not only air, but additionally moisture penetrates in to the tube, hence the sensor in the probe and the electronics around it is usually irreparably damaged. This may result in measurement errors and, in the worst case, even to failure of the particular level probe. To avoid any premature failure, the ingress of moisture into the ventilation tube must be completely prevented. Additional protection against moisture penetration through the ventilation tube is provided by fitting an air-permeable, but water-impermeable filter element at the end of the vent tube.
bare wires
Never to be ignored can be the transport of the liquid through high-humidity loads along the only limitedly protected internals of the cable, e.g. along the wires, completely right down to the submersible pressure transmitter. As a leading manufacturer, WIKA uses appropriate structural design to avoid fluid transport, so far as possible, in to the electronics of the submersible pressure transmitter. Due to molecular diffusion and capillary effects, a guaranteed one-hundred percent protection over the full duration of the submersible pressure transmitter, however, is never achievable.
Hence, it is recommended that the cable is always terminated in a waterproof junction box with the correct IP protection (e.g. IP65) which is matched to the installation location. If this cable junction box is subjected to weather and varying temperature conditions, additionally it is recommended to pay attention to a controlled pressure equalisation in order to avoid the formation of condensation or perspiration water and pumping effects. To handle this technical requirement, being an accessory to a submersible pressure transmitter, you’ll be able to order a connection box with an integrated air-permeable, water-impermeable membrane.
Ultimately, moisture ingress can occur not merely through the exposed end of the cable, but additionally through mechanical damage to the cable sheath or as a result of liquid diffusion due to improper chemical resistance of the cable material. In the article ?Selection criteria for preventing moisture-related failures of submersible pressure transmitters or level probes? this failure mode is described in detail.
WIKA offers comprehensive solutions for the hydrostatic-pressure level measurement. For further assistance in selecting the submersible pressure transmitter the most suitable for the application, please use our contact page.
Please find further information with this topic on our information platform ?Hydrostatic level measurement?

Leave a Comment